Integrated monitoring can detect critical events and improve alarm accuracy.

M J Navabi1, R C Watt, S T Hameroff, K C Mylrea.   

Abstract

A computer-based, integrated monitor system was designed and utilized to collect and interactively manage physiologic data (13 variables and 3 waveforms) from six routinely used operating room monitors. Various approaches were developed to reduce false alarms, classify waveforms, and recognize events. False alarms: false alarms in ECG heart rate detection were reduced from 37.3% to 2.6% (p=0.005) of total alarms using multi-variable analysis and rate-of-change limits. Waveform classification: using artificial neural networks (AN), CO2 waveforms were classified into (a) spontaneous, (b) mechanical, and (c) mechanical/with spontaneous breathing attempts. The system properly classified 47 of 71 spontaneous, 65 of 67 mechanical, and 37 of 44 mechanical breaths/with spontaneous breathing attempts. Another ANN was used for detection of elevated and depressed ST segments in the ECG signal. All ST segment elevations and depressions of 0.1 mV were correctly identified. Event recognition: an algorithm developed to identify endotracheal intubation correctly recognized 13 of 17 intubations. This resulted in a 42% reduction in low end-tidal-CO2 false alarms.

Entities:  

Mesh:

Year:  1991        PMID: 10114048     DOI: 10.1097/00004669-199107000-00001

Source DB:  PubMed          Journal:  J Clin Eng        ISSN: 0363-8855


  2 in total

1.  An expert system for monitor alarm integration.

Authors:  C Oberli; J Urzua; C Saez; M Guarini; A Ciprianio; B Garayar; G Lema; R Canessa; C Sacco; M Irarrazaval
Journal:  J Clin Monit Comput       Date:  1999-01       Impact factor: 2.502

2.  Algorithm to identify components of arterial blood pressure signals during use of an intra-aortic balloon pump.

Authors:  Z F Elghazzawi; J P Welch; Z Ladin; P Ford-Carleton; J B Cooper
Journal:  J Clin Monit       Date:  1993-09
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.