Literature DB >> 10101839

Characterisation of a thermophilic L-glutamate dehydrogenase biosensor for amperometric determination of L-glutamate by flow injection analysis.

N Pasco1, C Jeffries, Q Davies, A J Downard, A D Roddick-Lanzilotta, L Gorton.   

Abstract

Carbon paste wax electrodes incorporating thermophilic L-glutamate dehydrogenase, NADP and a polymeric toluidine blue O (poly-TBO) mediator have been characterised for the amperometric determination of L-glutamate at 313-318 K in a flow injection analysis (FIA) system. The biosensors exhibit good sensitivity, mechanical stability and reproducibilty, unlike carbon paste- or carbon wax-based electrodes under the same conditions. The carbon paste wax electrode responds linearly to L-glutamate up to 40 mM, the detection limit is 0.3 mM and the RSD (n = 10) for 5 mM L-glutamate was 7.6%. The response to some potential interferents has been quantified. Addition of finely ground hexaammineruthenium (III) trichloride ([Ru(NH3)6]Cl3) to the carbon paste wax electrodes decreases the FIA peak width and increases the peak current. The metal complex appears to accelerate the rate of oxidation of NAD(P)H by poly-TBO.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10101839     DOI: 10.1016/s0956-5663(98)00120-1

Source DB:  PubMed          Journal:  Biosens Bioelectron        ISSN: 0956-5663            Impact factor:   10.618


  2 in total

1.  RECENT DEVELOPMENTS IN ELECTROCHEMICAL SENSORS FOR THE DETECTION OF NEUROTRANSMITTERS FOR APPLICATIONS IN BIOMEDICINE.

Authors:  Rıfat Emrah Özel; Akhtar Hayat; Silvana Andreescu
Journal:  Anal Lett       Date:  2014-12-31       Impact factor: 2.329

2.  Label Free Detection of L-Glutamate Using Microfluidic Based Thermal Biosensor.

Authors:  Varun Lingaiah Kopparthy; Siva Mahesh Tangutooru; Eric J Guilbeau
Journal:  Bioengineering (Basel)       Date:  2015-01-12
  2 in total