N Pasco1, C Jeffries, Q Davies, A J Downard, A D Roddick-Lanzilotta, L Gorton.
Abstract
Carbon paste wax electrodes incorporating thermophilic L-glutamate dehydrogenase, NADP and a polymeric toluidine blue O (poly-TBO) mediator have been characterised for the amperometric determination of L-glutamate at 313-318 K in a flow injection analysis (FIA) system. The biosensors exhibit good sensitivity, mechanical stability and reproducibilty, unlike carbon paste- or carbon wax-based electrodes under the same conditions. The carbon paste wax electrode responds linearly to L-glutamate up to 40 mM, the detection limit is 0.3 mM and the RSD (n = 10) for 5 mM L-glutamate was 7.6%. The response to some potential interferents has been quantified. Addition of finely ground hexaammineruthenium (III) trichloride ([Ru(NH3)6]Cl3) to the carbon paste wax electrodes decreases the FIA peak width and increases the peak current. The metal complex appears to accelerate the rate of oxidation of NAD(P)H by poly-TBO.Entities:
Mesh:
Substances:
Year: 1999 PMID: 10101839 DOI: 10.1016/s0956-5663(98)00120-1
Source DB: PubMed Journal: Biosens Bioelectron ISSN: 0956-5663 Impact factor: 10.618