| Literature DB >> 10100160 |
P G Archer1, J M Balter, D A Ross, J A Hayman, H M Sandler.
Abstract
Over a period of approximately 3 years, our institution has implemented and refined a system of Stereotactic Radiosurgery (SRS) which utilizes the standard multi leaf collimator (MLC) of the Scanditronix MM50 Racetrack Microtron and treats in an arrangement of segmental "pseudo-arcs." This system employs a commercial BRW based stereotactic frame which is mounted to the treatment table. With the exception of the table-mounted frame hardware there have been no modifications to the treatment machine to accommodate these treatments. By use of standard evaluation parameters (e.g., treatment time, planning time, dose conformance and dose heterogeneity ratios) this system compares quite favorably with reported data from institutions treating SRS with either a GammaKnife or a standard linear accelerator with tertiary collimators.Entities:
Mesh:
Year: 1999 PMID: 10100160 DOI: 10.1016/s0958-3947(98)00048-x
Source DB: PubMed Journal: Med Dosim ISSN: 1873-4022 Impact factor: 1.482