Abstract

Failures of Gamma nails which treat unstable femoral fractures have been reported. In this paper, a finite element model to include a Gamma nail within a fractured femur was used to investigate the stresses in the Gamma nail. The effects for different types of fracture were investigated. The results show that its use for subtrochanteric fractures will cause higher stresses at the lag screw and upper distal screw insertion holes in the nail than when used for femur neck fractures.

Entities:  

Mesh:

Year:  1998        PMID: 10098612     DOI: 10.1016/s1350-4533(98)00079-4

Source DB:  PubMed          Journal:  Med Eng Phys        ISSN: 1350-4533            Impact factor:   2.242


  3 in total

1.  Breakage of intramedullary femoral nailing or femoral plating: how to prevent implant failure.

Authors:  Henrik C Bäcker; Mark Heyland; Chia H Wu; Carsten Perka; Ulrich Stöckle; Karl F Braun
Journal:  Eur J Med Res       Date:  2022-01-13       Impact factor: 2.175

2.  Comparison of Three Fixation Methods for Femoral Neck Fracture in Young Adults: Experimental and Numerical Investigations.

Authors:  Shabnam Samsami; Sadegh Saberi; Sanambar Sadighi; Gholamreza Rouhi
Journal:  J Med Biol Eng       Date:  2015-10-01       Impact factor: 1.553

3.  Strength⁻Durability Correlation of Osteosynthesis Devices Made by 3D Layer Manufacturing.

Authors:  Yoshimitsu Okazaki; Emiko Gotoh; Jun Mori
Journal:  Materials (Basel)       Date:  2019-01-31       Impact factor: 3.623

  3 in total