Efficacy and safety of the initial use of stability and onset criteria in implantable cardioverter defibrillators.

M Weber1, D Böcker, D Bänsch, J Brunn, M Castrucci, R Gradaus, G Breithardt, M Block.   

Abstract

INTRODUCTION: Inappropriate therapies are the most frequent adverse event in patients with implantable cardioverter defibrillators (ICDs). Most ICDs offer a stability criterion to discriminate ventricular tachycardia (VT) from atrial fibrillation and an onset criterion to discriminate VT from sinus tachycardia. The efficacy and safety of these criteria, if used immediately after implantation, is unknown. METHODS AND
RESULTS: In a case control study, 87 patients in whom stability and onset criteria had been activated immediately after ICD implantation were matched to 87 patients in whom these criteria had not been activated. The groups were matched for known predictors of inappropriate therapies. With stability and onset criteria off, 24 patients (28%) received inappropriate therapies due to atrial fibrillation (n = 14) or sinus tachycardia (n = 11); with stability and onset on, only 11 patients (13%) were treated by the ICD due to atrial fibrillation (n = 5) or sinus tachycardia (n = 7) (log rank: P = 0.029). Five patients suffered inappropriate therapies despite the fact that onset (n = 4) or stability (n = 1) criteria were not fulfilled once tachycardias continued for a prespecified duration. Only one patient experienced a failure to detect VT due to the onset criterion; none because of stability.
CONCLUSION: The immediate use of stability and onset criteria after ICD implantation reduces inappropriate therapies due to atrial fibrillation and sinus tachycardia. Because of the potential for underdetection of VT, this approach should be limited to tachycardia rates hemodynamically tolerated by the patient.

Entities:  

Mesh:

Year:  1999        PMID: 10090217     DOI: 10.1111/j.1540-8167.1999.tb00655.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  5 in total

Review 1.  Implantable dual-chamber cardioverter-defibrillator-pacemaker.

Authors:  D Pfeiffer; M Mende; A Hagendorff
Journal:  Curr Cardiol Rep       Date:  2000-07       Impact factor: 2.931

Review 2.  Methods of minimizing inappropriate implantable cardioverter-defibrillator shocks.

Authors:  B Schaer; S Osswald
Journal:  Curr Cardiol Rep       Date:  2000-07       Impact factor: 2.931

Review 3.  Algorithms for better arrhythmia discrimination in implantable cardioverter defibrillators.

Authors:  R Schimpf; C Wolpert; B Lüderitz
Journal:  Curr Cardiol Rep       Date:  2001-11       Impact factor: 2.931

4.  2015 HRS/EHRA/APHRS/SOLAECE expert consensus statement on optimal implantable cardioverter-defibrillator programming and testing.

Authors:  Bruce L Wilkoff; Laurent Fauchier; Martin K Stiles; Carlos A Morillo; Sana M Al-Khatib; Jesœs Almendral; Luis Aguinaga; Ronald D Berger; Alejandro Cuesta; James P Daubert; Sergio Dubner; Kenneth A Ellenbogen; N A Mark Estes; Guilherme Fenelon; Fermin C Garcia; Maurizio Gasparini; David E Haines; Jeff S Healey; Jodie L Hurtwitz; Roberto Keegan; Christof Kolb; Karl-Heinz Kuck; Germanas Marinskis; Martino Martinelli; Mark McGuire; Luis G Molina; Ken Okumura; Alessandro Proclemer; Andrea M Russo; Jagmeet P Singh; Charles D Swerdlow; Wee Siong Teo; William Uribe; Sami Viskin; Chun-Chieh Wang; Shu Zhang
Journal:  J Arrhythm       Date:  2016-02-01

5.  Standardized programming to reduce the burden of inappropriate therapies in implantable cardioverter defibrillators - Single centre follow up results.

Authors:  U Boles; E E Gul; L Fitzgerald; F Sadiq Ali; C Nolan; K Aldworth-Gaumond; D R Redfearn; A Baranchuk; B Glover; C Simpson; H Abdollah; K A Michael
Journal:  Indian Pacing Electrophysiol J       Date:  2017-10-27
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.