Literature DB >> 10080090

Numerical model of fluid flow and oxygen transport in a radial-flow microchannel containing hepatocytes.

G A Ledezma1, A Folch, S N Bhatia, U J Balis, M L Yarmush, M Toner.   

Abstract

The incorporation of monolayers of cultured hepatocytes into an extracorporeal perfusion system has become a promising approach for the development of a temporary bioartificial liver (BAL) support system. In this paper we present a numerical investigation of the oxygen tension, shear stress, and pressure drop in a bioreactor for a BAL composed of plasma-perfused chambers containing monolayers of porcine hepatocytes. The chambers consist of microfabricated parallel disks with center-to-edge radial flow. The oxygen uptake rate (OUR), measured in vitro for porcine hepatocytes, was curve-fitted using Michaelis-Menten kinetics for simulation of the oxygen concentration profile. The effect of different parameters that may influence the oxygen transport inside the chambers, such as the plasma flow rate, the chamber height, the initial oxygen tension in the perfused plasma, the OUR, and K(m) was investigated. We found that both the plasma flow rate and the initial oxygen tension may have an important effect upon oxygen transport. Increasing the flow rate and/or the inlet oxygen tension resulted in improved oxygen transport to cells in the radial-flow microchannels, and allowed significantly greater diameter reactor without oxygen limitation to the hepatocytes. In the range investigated in this paper (10 microns < H < 100 microns), and for a constant plasma flow rate, the chamber height, H, had a negligible effect on the oxygen transport to hepatocytes. On the contrary, it strongly affected the mechanical stress on the cells that is also crucial for the successful design of the BAL reactors. A twofold decrease in chamber height from 50 to 25 microns produced approximately a fivefold increase in maximal shear stress at the inlet of the reactor from 2 to 10 dyn/cm2. Further decrease in chamber height resulted in shear stress values that are physiologically unrealistic. Therefore, the channel height needs to be carefully chosen in a BAL design to avoid deleterious hydrodynamic effects on hepatocytes.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10080090     DOI: 10.1115/1.2798043

Source DB:  PubMed          Journal:  J Biomech Eng        ISSN: 0148-0731            Impact factor:   2.097


  11 in total

1.  Gray-scale photolithography using microfluidic photomasks.

Authors:  Chihchen Chen; Danny Hirdes; Albert Folch
Journal:  Proc Natl Acad Sci U S A       Date:  2003-02-06       Impact factor: 11.205

2.  Computerized microfluidic cell culture using elastomeric channels and Braille displays.

Authors:  Wei Gu; Xiaoyue Zhu; Nobuyuki Futai; Brenda S Cho; Shuichi Takayama
Journal:  Proc Natl Acad Sci U S A       Date:  2004-10-28       Impact factor: 11.205

3.  Metabolic response of perfused livers to various oxygenation conditions.

Authors:  Mehmet A Orman; Marianthi G Ierapetritou; Ioannis P Androulakis; Francois Berthiaume
Journal:  Biotechnol Bioeng       Date:  2011-08-04       Impact factor: 4.530

Review 4.  New tools and new biology: recent miniaturized systems for molecular and cellular biology.

Authors:  Morgan Hamon; Jong Wook Hong
Journal:  Mol Cells       Date:  2013-12-02       Impact factor: 5.034

Review 5.  Challenges and Opportunities in the Design of Liver-on-Chip Microdevices.

Authors:  Avner Ehrlich; Daniel Duche; Gladys Ouedraogo; Yaakov Nahmias
Journal:  Annu Rev Biomed Eng       Date:  2019-06-04       Impact factor: 9.590

6.  Three-dimensional numerical modeling and computational fluid dynamics simulations to analyze and improve oxygen availability in the AMC bioartificial liver.

Authors:  Guy Mareels; Paul P C Poyck; Sunny Eloot; Robert A F M Chamuleau; Pascal R Verdonck
Journal:  Ann Biomed Eng       Date:  2006-10-10       Impact factor: 3.934

Review 7.  Is there a future for liver-assist devices?

Authors:  P Rosenthal
Journal:  Curr Gastroenterol Rep       Date:  2000-02

8.  A microfluidic hepatic coculture platform for cell-based drug metabolism studies.

Authors:  Eric Novik; Timothy J Maguire; Piyun Chao; K C Cheng; Martin L Yarmush
Journal:  Biochem Pharmacol       Date:  2009-11-27       Impact factor: 5.858

9.  Effect of oxygen concentration on viability and metabolism in a fluidized-bed bioartificial liver using ³¹P and ¹³C NMR spectroscopy.

Authors:  Rex E Jeffries; Michael P Gamcsik; Kayvan R Keshari; Peter Pediaditakis; Andrey P Tikunov; Gregory B Young; Haakil Lee; Paul B Watkins; Jeffrey M Macdonald
Journal:  Tissue Eng Part C Methods       Date:  2012-09-28       Impact factor: 3.056

Review 10.  Design and application of microfluidic systems for in vitro pharmacokinetic evaluation of drug candidates.

Authors:  T J Maguire; E Novik; P Chao; J Barminko; Y Nahmias; M L Yarmush; K-C Cheng
Journal:  Curr Drug Metab       Date:  2009-12       Impact factor: 3.731

View more