Evaluation of the impedance technique for cryosurgery in a theoretical model of the head.

M M Radai1, S Abboud, B Rubinsky.   

Abstract

Bioimpedance is a noninvasive technique that produces information on the electrical characteristics of tissue inside the body from currents injected and electrical potentials measured on the surface of the body. Because freezing causes a large increase in tissue electrical impedance we thought that it may also cause significant changes in the surface electrical potential making the bioimpedance technique suitable for noninvasive monitoring and imaging of cryosurgery. To evaluate the feasibility of the bioimpedance technique in cryosurgery we examined, as a case study, a theoretical model for the electrical potentials during brain cryosurgery. A three-dimensional spherical model was used to calculate the change in the electrical potential distribution in the head as a function of the current source location and the size of the frozen lesion in the brain. The numerical calculations were executed using the finite volume method and the iterative successive over relaxation method. The results demonstrate that, indeed, freezing inside the head produces measurable changes in the electrical potential on the outer surface-the scalp. Copyright 1999 Academic Press.

Mesh:

Year:  1999        PMID: 10079129     DOI: 10.1006/cryo.1998.2148

Source DB:  PubMed          Journal:  Cryobiology        ISSN: 0011-2240            Impact factor:   2.487


  2 in total

1.  Frequency-division multiplexing for electrical impedance tomography in biomedical applications.

Authors:  Yair Granot; Antoni Ivorra; Boris Rubinsky
Journal:  Int J Biomed Imaging       Date:  2007

2.  Determination of single cryoablation outcome within 30 to 60 seconds of freezing based on ice impedance.

Authors:  Boaz Avitall; Ken S Lizama; Arthur Kalinski; Nicolas Coulombe; Timothy G Laske
Journal:  J Cardiovasc Electrophysiol       Date:  2019-08-19
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.