Physics of rotating gamma systems for stereotactic radiosurgery.

S J Goetsch1, B D Murphy, R Schmidt, J Micka, L De Werd, Y Chen, S Shockley.   

Abstract

PURPOSE: The purpose of this study was to evaluate the characteristics of a new rotating gamma system for stereotactic radiosurgery by comparison with a well accepted system. METHODS AND MATERIALS: A novel gamma unit for stereotactic radiosurgery has been developed and distributed to 15 hospitals in China. The unit contains 30 cobalt-60 gamma radiation sources with initial activity of 200 Ci (7.4 x 10(12) Bq) each. The sources are positioned along 30 arcs, and rotate continuously as a group in an axis orthogonal to the patient's body. Measurements have been made on a representative unit installed in the Auhai Radiosurgery Center at the Beijing Navy General Hospital in the People's Republic of China. Ionization chambers calibrated by an American accredited dosimetry calibration laboratory were used for these measurements, as well as radiochromic film and thermoluminescent dosimeters. The unit tested utilizes collimators of nominal diameters of 4, 8, 14, and 18 mm. Radiochromic film samples from a Leksell Model U Gamma Knife were evaluated by the same laboratory and are presented for comparison. The treatment planning system was not evaluated.
RESULTS: Radiation-absorbed dose rates and profiles measured for this unit are comparable to those previously measured with the same techniques for the Leksell Model U Gamma Knife units in San Diego and Atlanta.
CONCLUSION: This unit is capable of producing well collimated beams of high energy photons, suitable for stereotactic radiosurgery. It has similar physical characteristics to those previously reported for the Leksell Model U Gamma Knife unit.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10078657     DOI: 10.1016/s0360-3016(98)00428-3

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  5 in total

1.  Image-guided therapy: evolution and breakthrough.

Authors:  Pascal Haigron; Jean-Louis Dillenseger; Limin Luo; Jean-Louis Coatrieux
Journal:  IEEE Eng Med Biol Mag       Date:  2010 Jan-Feb

2.  Dosimetry of a novel rotating gamma system for stereotactic radiosurgery.

Authors:  Kuang-Ming Yang; Steen J Madsen; Wei-Li Chen; Wei-Peng Kuan; Ronald F Young; Ching Chen
Journal:  J Radiosurg SBRT       Date:  2011

3.  Rotating Gamma System Irradiation: A Promising Treatment for Low-grade Brainstem Gliomas.

Authors:  Pham Cam Phuong; Nguyen Quang Hung; Tran Bao Ngoc; Dirk Rades; Mai Trong Khoa
Journal:  In Vivo       Date:  2017 Sep-Oct       Impact factor: 2.155

4.  Long-term outcomes of rotating gamma knife for vestibular schwannoma: A 4-year prospective longitudinal study of 89 consecutive patients in Vietnam.

Authors:  Hung Dinh Kieu; Duong Ngoc Vuong; Khoa Trong Mai; Phuong Cam Pham; Tam Duc Le
Journal:  Surg Neurol Int       Date:  2021-11-30

5.  Dosimetry and Treatment Descriptions Using the First Completely Automated Stereotactic Intracranial Radiosurgery Rotating Gamma Ray Unit in America.

Authors:  Eduardo E Lovo; Fidel J Campos; Victor E Caceros; Mario H Minervini; William A Reyes
Journal:  Cureus       Date:  2018-04-27
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.