Ionization chamber, electrometer, linear accelerator, field size, and energy dependence of the polarity effect in electron dosimetry.

C R Ramsey1, K M Spencer, A L Oliver.   

Abstract

Plane-parallel ionization chambers are the instrument of choice for use in electron calibration and dosimetry, but these chambers may exhibit large polarity effects. This study concentrates on measuring the dependence of the polarity error at various mean energies using different linear accelerator, field size, ion chamber, and electrometer combinations. The polarity error was shown to increase for all four ionization chambers as the mean energy at depth decreased, but was always less than one percent at d(max). Polarity error dependence was also observed for similar plane-parallel chambers, varying field sizes, and different linear accelerators, but no polarity error dependence was observed for similar cylindrical chambers and different electrometers. Measurements of the polarity error can be used to develop correction factors for future measurements that will help minimize the time spent performing electron dosimetry and calibrations. These correction factors can be used to calculate the correct reading without the need to reverse the chamber bias, thus reducing the number of measurements required.

Entities:  

Mesh:

Year:  1999        PMID: 10076977     DOI: 10.1118/1.598507

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  2 in total

1.  Polarity correction factor for flattening filter free photon beams in several cylindrical ionization chambers.

Authors:  Toshiyuki Ogata; Kazuyuki Uehara; Masao Nakayama; Shinji Tsudou; Takashi Masutani; Takanobu Okayama
Journal:  Radiol Phys Technol       Date:  2016-02-12

2.  Percent depth-dose distribution discrepancies from very small volume ion chambers.

Authors:  Vikren Sarkar; Brian Wang; Hui Zhao; Bart Lynch; Joshua A James; Kiernan T McCullough; Bill J Salter
Journal:  J Appl Clin Med Phys       Date:  2015-03-08       Impact factor: 2.102

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.