Haemolysis during cardiopulmonary bypass: an in vivo comparison of standard roller pumps, nonocclusive roller pumps and centrifugal pumps.

S D Hansbro1, D A Sharpe, R Catchpole, K R Welsh, C M Munsch, J P McGoldrick, P H Kay.   

Abstract

Cardiopulmonary bypass (CPB) involves the use of either an occlusive roller pump or centrifugal pump. Damage to blood elements, including haemolysis, may arise from occlusion when using a roller pump; the appropriate degree of occlusion has not yet been determined scientifically. Centrifugal and nonocclusive roller pumps are reputed to reduce haemolysis. The objective of this study was to compare haemolysis caused by a standard roller pump with a dynamically set nonocclusive roller pump and with a centrifugal pump. We prospectively randomized 60 patients undergoing routine coronary artery surgery into three groups: standard roller pump (STD, n = 20), dynamically set roller pump (DYN, n = 20), or centrifugal pump (CEN, n = 20). The level of plasma free haemoglobin (FHb) was measured preoperatively, and the rate of formation of FHb (in mg/dl/min) was determined at the end of the ischaemic phase and at the end of CPB. Cardiotomy suction blood was isolated for the ischaemic phase and returned before the end of CPB. It was found that there were no differences between the groups in demographic or operative variables. The rate of formation of FHb at the end of the ischaemic phase was similar for all groups (STD 0.108 +/- 0.10, DYN 0.117 +/- 0.08, CEN 0.129 +/- 0.07). At the end of CPB, after return of the cardiotomy suction blood, there was a significant (< 0.001) increase in the rate of formation of FHb in all groups. The increase was similar for each of the groups (STD 0.424 +/- 0.17, DYN 0.481 +/- 0.20, CEN 0.471 +/- 0.18). We conclude that the rates of haemolysis are similar for each of the pump types, and no benefit is conferred by the use of either a dynamically set roller pump or a centrifugal pump compared with the standard roller pump. The return of the cardiotomy suction blood to the circulation is the principal source of plasma free haemoglobin.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10074641     DOI: 10.1177/026765919901400102

Source DB:  PubMed          Journal:  Perfusion        ISSN: 0267-6591            Impact factor:   1.972


  3 in total

1.  Influence of transfusion technique on survival of autologous red blood cells in the dog.

Authors:  Ruth I McDevitt; Craig G Ruaux; Wendy I Baltzer
Journal:  J Vet Emerg Crit Care (San Antonio)       Date:  2011-05-10

2.  Getting it right: optimizing the patient and technique for the procedure.

Authors:  Alfred H Stammers
Journal:  J Extra Corpor Technol       Date:  2009-12

Review 3.  Hemolysis in cardiac surgery patients undergoing cardiopulmonary bypass: a review in search of a treatment algorithm.

Authors:  Leen Vercaemst
Journal:  J Extra Corpor Technol       Date:  2008-12
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.