C C Ginocchio1, S Tetali, D Washburn, F Zhang, M H Kaplan.
Abstract
This study compared levels of human immunodeficiency virus type 1 RNA in plasma as measured by the Quantiplex branched-DNA and NucliSens nucleic acid sequence-based amplification assays. RNA was detectable in 118 of 184 samples (64.13%) by the Quantiplex assay and in 171 of 184 samples (92.94%) by the NucliSens assay. Regression analysis indicated that a linear relationship existed between the two sets of values (P < 0.0001), although the Quantiplex and NucliSens values were significantly different (P < 0.001), with the NucliSens values being approximately 0.323 log higher. Spearman correlation analysis indicated that the overall changes in patient viral load patterns were highly correlative between the two assays: r = 0.912, P < 0.0001. The lower limits of sensitivity were determined to be approximately 100 copies/ml and 1,200 to 1,400 copies/ml for the NucliSens and Quantiplex assays, respectively.Entities:
Mesh:
Substances:
Year: 1999 PMID: 10074556 PMCID: PMC88679
Source DB: PubMed Journal: J Clin Microbiol ISSN: 0095-1137 Impact factor: 5.948