Abstract
A method is described for the qualitative and semi-quantitative simultaneous determination of both non-macrocyclic and macrocyclic trichothecene biotoxins from samples derived from indoor environments. The method includes extraction, sample pre-treatment and reversed-phase HPLC separation followed by tandem mass spectrometric identification and quantification using electrospray ionization on a quadrupole ion trap mass analyser. Aqueous methanol was used in the initial extraction and solvent partitioning and solid-phase extraction in the purification of samples. The HPLC separation was run on-line with electrospray ionization MS-MS detection. The detection limits and recoveries of the procedure varied from 1 to 1000 pg and from 31 to 92%, respectively. As the method includes few and not very labour intensive sample treatment steps, it should allow for a high throughput of samples with good prospects of automation.Entities:
Mesh:
Substances:
Year: 1998 PMID: 10071380 DOI: 10.1039/a803813i
Source DB: PubMed Journal: Analyst ISSN: 0003-2654 Impact factor: 4.616