Dose errors in the near field of an HDR brachytherapy stepping source.

T Wong1, S Wallace, W Fernando, W Schumer, G Quong.   

Abstract

The dose rate at point P at 0.25 cm in water from the transverse bisector of a straight catheter with an active stepping source (Nucletron microSelectron HDR source) with a dwell length of 2 cm was calculated using Monte Carlo code MCNP 4.A. The source step sizes were 1 cm and 0.25 cm. The Monte Carlo (MC) results were used for comparison with the results calculated with the Nucletron brachytherapy planning system (BPS) formalism, first with BPS variants and then with its respective MC calculated radial dose function and anisotropy function. The dose differences at point P calculated using the BPS formalism and variants are +15.4% and +3.1% for the source step size of 1 cm and 0.25 cm respectively. This reduction in dose difference is caused by the increased importance of errors in the anisotropy function with the smaller step size, which counter the errors in the radial dose function. Using the MC calculated radial dose function and anisotropy function with the BPS formalism. 1% dose calculation accuracy can be achieved, even in the near field, with negligible extra demand on computation time.

Entities:  

Mesh:

Year:  1999        PMID: 10070787     DOI: 10.1088/0031-9155/44/2/005

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  4 in total

1.  Monte Carlo dose calculation of GZP6 (60)Co stepping source based on a matrix shift technique.

Authors:  Mohammad Taghi Bahreyni Toossi; Malihe Abdollahi; Mahdi Ghorbani
Journal:  Rep Pract Oncol Radiother       Date:  2010-12-21

2.  A Monte Carlo study on dose distribution validation of GZP6 (60)Co stepping source.

Authors:  Mohammad Taghi Bahreyni Toossi; Maliheh Abdollahi; Mahdi Ghorbani
Journal:  Rep Pract Oncol Radiother       Date:  2012-11-21

3.  Dosimetric parameters of three new solid core I-125 brachytherapy sources.

Authors:  Timothy D Solberg; John J DeMarco; Geoffrey Hugo; Robert E Wallace
Journal:  J Appl Clin Med Phys       Date:  2002       Impact factor: 2.102

4.  Intensity Modulated High Dose Rate (HDR) Brachytherapy Using Patient Specific 3D Metal Printed Applicators: Proof of Concept.

Authors:  James J Sohn; Mitchell Polizzi; Sang-Won Kang; Woo-Hyeong Ko; Yong-Hyun Cho; Keun-Yong Eom; Jin-Beom Chung
Journal:  Front Oncol       Date:  2022-02-10       Impact factor: 5.738

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.