Validation of haemodialysis recirculation and access blood flow measured by thermodilution.

D Schneditz1, E Wang, N W Levin.   

Abstract

BACKGROUND: Recirculation (R) and access blood flow (Qac) measurements are considered useful indicators of adequate delivery of haemodialysis. It was the purpose of this study to compare measurements of R and Qac obtained by two different techniques which are based on the same principle of indicator dilution, but which differ because of the characteristics of the injection and detection of the different indicators used.
METHODS: Recirculation measured by a thermal dilution technique using temperature sensors (BTM, Fresenius Medical Care) was compared with recirculation measured by a validated saline dilution technique using ultrasonic transducers placed on arterial and venous segments of the extracorporeal circulation (HDM, Transonic Systems, Inc.). Calculated access flows were compared by Bland Altman analysis. Data are given as mean +/- SD.
RESULTS: A total of 104 measurements obtained in 52 treatments (17 patients, 18 accesses) were compared. Recirculation measured with correct placement of blood lines and corrected for the effect of cardiopulmonary recirculation using the 'double recirculation technique' was -0.02 +/- 0.14% by the BTM technique and not different from the 0% measured by the HDM technique. Recirculation measured with reversed placement of blood lines and corrected for the effect of cardiopulmonary recirculation was 19.66 +/- 10.77% measured by the BTM technique compared with 20.87 +/- 11.64% measured by the HDM technique. The difference between techniques was small (-1.21 +/- 2.44%) albeit significant. Access flow calculated from BTM recirculation was 1328 +/- 627 ml/min compared with 1390 +/- 657 ml/min calculated by the HDM technique. There was no bias between techniques.
CONCLUSION: BTM thermodilution yields results which are consistent with the HDM ultrasound dilution technique with regard to both recirculation and access flow measurement.

Entities:  

Mesh:

Year:  1999        PMID: 10069192     DOI: 10.1093/ndt/14.2.376

Source DB:  PubMed          Journal:  Nephrol Dial Transplant        ISSN: 0931-0509            Impact factor:   5.992


  3 in total

1.  Ultrasound dilution and thermodilution versus color Doppler ultrasound for arteriovenous fistula assessment in children on hemodialysis.

Authors:  Vasiliki Karava; Theresa Kwon; Gilbert Franco; Deschenes Georges; Marie-Alice Macher; Julien Hogan
Journal:  Pediatr Nephrol       Date:  2019-07-19       Impact factor: 3.714

2.  A model to calculate cardiac output in hemodialysis patients by thermodilution.

Authors:  Ahmed Alayoud; Kawtar Hassani; Mohammed Benyahia
Journal:  Theor Biol Med Model       Date:  2012-06-21       Impact factor: 2.432

3.  Prediction of vascular access stenosis: Blood temperature monitoring with the Twister versus static intra-access pressure ratio.

Authors:  Yoo Jin Choi; Young-Ki Lee; Hayne Cho Park; Eun Yi Kim; Ajin Cho; Chaehoon Han; Sun Ryoung Choi; Hanmyun Kim; Eun-Jung Kim; Jong-Woo Yoon; Jung-Woo Noh
Journal:  PLoS One       Date:  2018-10-29       Impact factor: 3.240

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.