| Literature DB >> 10036138 |
Y Liu1, W Kati, C M Chen, R Tripathi, A Molla, W Kohlbrenner.
Abstract
A general method is presented here for the determination of the Km, kcat, and kcat/Km of fluorescence resonance energy transfer (FRET) substrates using a fluorescence plate reader. A simple empirical method for correcting for the inner filter effect is shown to enable accurate and undistorted measurements of these very important kinetic parameters. Inner filter effect corrected rates of hydrolysis of a FRET peptide substrate by hepatitis C virus (HCV) NS3 protease at various substrate concentrations enabled measurement of a Km value of 4.4 +/- 0.3 microM and kcat/Km value of 96,500 +/- 5800 M-1 s-1. These values are very close to the HPLC-determined Km value of 4.6 +/- 0.7 microM and kcat/Km value of 92,600 +/- 14,000 M-1 s-1. We demonstrate that the inner filter effect correction of microtiter plate reader velocities enables rapid measurement of Ki and Ki' values and kinetic inhibition mechanisms for HCV NS3 protease inhibitors. Copyright 1999 Academic Press.Entities:
Mesh:
Substances:
Year: 1999 PMID: 10036138 DOI: 10.1006/abio.1998.3014
Source DB: PubMed Journal: Anal Biochem ISSN: 0003-2697 Impact factor: 3.365