Precision and accuracy of an electrogoniometer.

H W Christensen1.   

Abstract

BACKGROUND: Spinal range of motion has been studied by use of a variety of different methods. One electrogoniometer has been used as a noninvasive, reliable, and accurate method. The accuracy and precision of this device has not been sufficiently documented.
OBJECTIVE: To evaluate the precision and accuracy of an electrogoniometer in comparison with two manual protractors. STUDY
DESIGN: This study investigated the precision and accuracy of an electrogoniometer (OSI CA 6000).
METHODS: The precision and accuracy of the electrogoniometer were determined by a series of tests against two manual protractors. Six different movement directions were tested and 4 or 5 different angles for a given movement direction were evaluated.
RESULTS: The precision of the electrogoniometer was found to be +/- 0.1 degree for 4 different angles (10 degrees, 20 degrees, 40 degrees, and 60 degrees) in 6 different movement directions (flexion, extension, left lateral flexion, right lateral flexion, left rotation, and right rotation). The accuracy was tested by comparing both manual protractors with the electrogoniometer. Substantial disagreement for 27 of 28 datasets was found. In only one dataset was there agreement between the two manual protractors and the electrogoniometer. The mean differences were in the interval 2.0% to 11.5% between the electrogoniometer and the protractors.
CONCLUSION: The electrogoniometer evaluated has a very high precision when range of motion measurements are made. However, the accuracy of the device is less than acceptable.

Entities:  

Mesh:

Year:  1999        PMID: 10029943     DOI: 10.1016/s0161-4754(99)70099-0

Source DB:  PubMed          Journal:  J Manipulative Physiol Ther        ISSN: 0161-4754            Impact factor:   1.437


  5 in total

Review 1.  Working postures: a literature review.

Authors:  Edgar Ramos Vieira; Shrawan Kumar
Journal:  J Occup Rehabil       Date:  2004-06

2.  Charge couple device-based system for 3-dimensional real time positioning on the assessment of segmental range of motion of lumbar spine.

Authors:  Ping Zhao; Li-jun Chen; Jing Guan; Li Pan; Hui Ding; Hai-shu Ding
Journal:  Chin J Integr Med       Date:  2005-12       Impact factor: 1.978

3.  Effectiveness of adaptive silverware on range of motion of the hand.

Authors:  Susan S McDonald; David Levine; Jim Richards; Lauren Aguilar
Journal:  PeerJ       Date:  2016-02-15       Impact factor: 2.984

4.  Asymmetry of spinal segments mobility in canoeists and its relationship with racing speed.

Authors:  Mateusz Rynkiewicz; Tadeusz Rynkiewicz; Włodzimierz Starosta
Journal:  J Hum Kinet       Date:  2013-03-28       Impact factor: 2.193

5.  Reliability and measurement error of frontal and horizontal 3D spinal motion parameters in 219 patients with chronic low back pain.

Authors:  Steen Harsted; Rune M Mieritz; Gert Bronfort; Jan Hartvigsen
Journal:  Chiropr Man Therap       Date:  2016-04-04
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.