Literature DB >> 1000800

Energy-dispersive x-ray fluorescence spectroscopy and inductively coupled plasma emission spectrometry evaluated for multielement analysis in complex biological matrices.

R D Irons, E A Schenk, R D Giauque.   

Abstract

Energy-dispersive x-ray spectroscopy and inductively coupled plasma emission spectrometry were evaluated as methods for routine multielement analysis of biological material. Standard samples included Standard Reference Materials (National Bureau of Standards), compounded mixtures, and supplements that provided a wide range of elemental concentrations for analysis. Elements included in this study were Zn, Pb, Ni, Mn, Fe, Mg, Cu, Ca, As, Se, Br, Rb, and Sr. Standards were analyzed as unknowns by participating laboratories. The two methods were evaluated for sensitivity, precision, and accuracy, and the results compared to those obtained for atomic absorption spectrometric analysis of identical standard unknowns. Both methods compared favorably and both were determined to be highly reliable for such an application. Advantages and disadvantages of each method are compared and discussed.

Entities:  

Mesh:

Substances:

Year:  1976        PMID: 1000800

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  1 in total

1.  Reliability of the ICP-AES for trace elements studies of biological materials.

Authors:  M A Chaudhri; P Hannaker
Journal:  Biol Trace Elem Res       Date:  1987-08       Impact factor: 3.738

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.